2.1.3 Slide (DIP) Switch Settings
The slide (DIP) switch can be observed (Figure 1) when the back cover is removed for installation
or removal of the dosimeter. If changes need to be made to the slide switch, open the case on the
rear of the unit.
2.1.3.1 With the help of Figure 1, locate the slide switch.
2.1.3.2 Turn the appropriate switches to the ON or OFF position depending on the required
function. Table 1 provides a functional description of each switch.
2.1.3.3 Close the battery cover.
Table 1: PAM-TRX Slide Switch Settings
FUNCTION SWITCH ON OFF
RED LED
SW1 enabled disabled
BUZZER
SW2 enabled disabled
VIBRATOR
SW3 enabled disabled
DOSIMETER
Power OUT
SW4 3.3 VDC to J1-Ring (For
use with EAM or PEA-
100)
None
DC Alarm Signal
OUT
SW5 DC Alarm signal to J1 –
Ring (LED or earphone)
NO
Magnetic Test
Reed Switch
SW6 Enabled disabled
ALARM Delay
SW7 YES NO
ALARM Latch
SW8 YES NO
BLUE LED
SW9 Receiver On/Mapping NO
GREEN LED
SW10 Flash on TXD disabled
CAUTION: SW4 and SW5 are exclusive, and only one must be turned on. Set to OFF
for mono earphone or LED plug use.
2.1.4 Connecting to an External Alarm Device
Test Button
Access
Reed Switch
location
(refer to Figure 1)
Battery
Cover