The publication, translation or reproduction, either party or wholly, of this document are not allowed without our written consent. 17
3. Operating Instructions
3.1 Preparation for Use
Remove the instrument from the shipping container and check for any physical damage. In the case of damage,
report it to the carrier and contact MGP Instruments.
Caution: Do not attempt to install or operate damaged equipment since safety and
pe
rformance may be affected.
3.2 Starting-up
3.2.1 Connect the antenna to the WRM2 PRSMA connector located on the top of the enclosure. The
antenna is only needed when transmitting data over-the- air.
3.2.2 Switch dipswitch 3, for data output from the WRM2 radio, to the ON position or dipswitch 5 &
6 to the ON position when programming the internal WRM2 radio See “Wireless Remote
Monitoring Operating Manual”, 15-00043 for configuring the WRM2 radio.
Note: Ensure dipswitches 1,2,3,7 and 8 are in the ON position when deploying
the unit for field use. Dipswitch 4 must be in the OFF position at on times
for field use and calibration.
3.2.3 Switch dipswitch 7&8 to the ON position to engage the battery backup
3.2.4 Connect the external power supply to the locking miniature power supply connector
3.2.5 Connect external WRM transmitter if used and ensure that the baud rate for the DRM and the
W
RM External Transmitter are set to 300 bps. The interval level for the DRM needs to be set to
ZERO (0), using RMV, so that the WRM transmitter can request the data from the device.
Note: By engaging the battery or the external power supply, the meter will start
operating. The unit is not supplied with an On-Off switch.
3.2.6 NOTE: (Firmware >= 071111 (MARCH 2008) and board updates) By pressing the ‘ACK
‘button once on the DRM-1 external display or the DRM-2 will show the current threshold rate
setting, thr.0 then the rate setting i.e., 300. By pressing the ACK button a second time and so
forth will show user preset threshold settings that were configured using the RMV software.
See section 4.2.10 for RMV software Dose Rate Threshold settings. See section 4.2.10 for more
information on setup and usage
A
CK thr.0 = selected ACK thr.1 ACK thr.2 ACK thr.3 ACK thr.4