MGP WRM2 Especificaciones Pagina 38

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 62
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 37
15-00044
Revision 5
April 2008
The publication, translation or reproduction, either party or wholly, of this document are not allowed without our written consent. 38
4. Click OK, the program enters the main window.
5. The caption in the main window (linearity report) displays the previous template. In the displayed
template the operator executes the required operations to complete the calibration and linearity.
6. To start the procedure, click File/New Report.
10.3 Template & Parameters Setup
The template includes setup parameters and instrument parameters. To select or create a new template, proceed
a
s follows.
To select an existing template
The RMC program is supplied with a number of default templates. Click File/ Open template and select the
required template.
To create a new template
a) Fill the following parameters in the displayed template: Src
(source), Measuring Point including units, Distance, and
Attenuators. To select between mR/h and R/h, or cps and
Kcps, or cpm and Kcpm, click each of the unit’s columns.
b) Select Parameters / Setup from the pull down menu.
Complete the following
Customer - Enter header (up to 4 lines) for the calibration &
linearity certificate.
Accepted accuracy:
As Left - Enter maximum allowed deviation
As Found - Enter maximum allowed deviation
Vista de pagina 37
1 2 ... 33 34 35 36 37 38 39 40 41 42 43 ... 61 62

Comentarios a estos manuales

Sin comentarios