10.6.1 Linearity Report Window Description
Src Type of source employed for calibration and linearity. The source type is defined in the set-up parameters.
Meas.Point - Radiation field where linearity tests are performed.
U
nit - Measurement units.
Distance and Att. - Distance between detector and source and attenuator factor.
As Found - Instrument reading prior to calibration.
As Left - Instrument reading after calibration.
Err% - Reading deviation % between measuring point (calculated value) and meter reading.
P/F - Notifies that linearity checking is correct.
Serial Number - Instrument number, may be changed in Parameters / Instrument.
Accepted accuracy - Permitted deviation in % between the instrument reading and radiation. Defined in the
Parameters / Set up.
Sources [Src] - Source type by which calibration and linearity were performed. Defined in the set-up
p
arameters.
Customer Header- Customer: Enter header (up to 4 lines) for the calibration & linearity certificate.
Curr. Calibrator - Present calibrator's name and calibration date.
Prev. Calibrator - Former calibrator's name and calibration date.
Factors - Factors' value.
(1) Checking adequate connection between instrument and computer. If after the fifth time there is no
communication, an alarm message is displayed.
(2) 0.00 mR/h - Instrument reading is updated every second.
(3) Closed - After checking the first reading by pressing X on the tested point line, it is recommended to perform
File
Save as. From now on, the file is automatically updated and saved after checking of each tested point. As
long as the Closed text box is not marked, it is possible to make changes while entering and exiting the file. At
the end of the linearity check process, marking in the Closed text box will make the file "read only". From now
on, when entering the saved file marked Closed, it will be impossible to perform any changes in the Linearity
Report window.