15-00044
Revision 5
April 2008
The publication, translation or reproduction, either party or wholly, of this document are not allowed without our written consent. 30
6.1 Set Meter Parameter via the PC
The parameters are set via the PC using the RMV software.
PC to meter
0 1 2 3 4 5 6 7 8 9 10
LF Status D5 D4 D3 D2 D1 D0 0F 0F CR
Status byte:
Read meter parameters=30H
Set Baud=31H
Set Time=34H
Set Threshold alarm value.=35H
Set Dose status = 37H
Set meter serial no. = 38H
6.1.1 Read the Meter Parameters=30H
PC to meter
0 1 2 3 4 5 6 7 8 9 10
LF 30 20 20 20 20 20 20 0F 0F CR
Meter to PC
0 1 2 3 4-9 10-14 15 16-19 20 21-26
LF 30 Meter
type
Meter
unit
Meter
#
Thr.
value
Meter
baud
Down
load
time
Dose
En/Dis
Firmware
version
Byte 2: Byte 3: byte 15: byte 20
AMP50 = 30 mR/h = 30 300=30 disable =30
AMP100=31 uSv/h = 31 4800=31 enable=31
AMP200=32 9600=32
19200=33
byte 10-14 : threshold X.YZW * 10 T (X=byte10, Y=11, Z=12, W=13, T=14)
27-28 29
36h,37h
CR