(6) New Factor: - The New Factor will be displayed and is a function of the actual current reading against the
Radiation field that was typed into the window. If the calculated factor value is lower than 0.6 or higher than
1.4, calibration will be not performed.
(7) Force 1 Select in case the operator wants to force the calibration factor to 1.00. Use this when an
incorrect factor was stored.
(
8) After the factor changing, press Update Factor. Ensure to obtain the status: Factor Accepted.
(9) Update Details - Press Update Details.
(10) Ensure to obtain Status: Details Accepted.
(11) End CalibrationClick to quit the calibration function and return to the main menu.
(12) Background sub (currently not in use).
10.6 Linearity
To enter the Linearity Report window press File New Report. The Linearity Report window includes
As Found and As Left columns. Before calibration and linearity check, perform the proceeding steps in the
following order:
1) Check the linearity in As Found column.
2) If linearity fail is obtained the instrument is defective and cannot be calibrated (P/F –fail).
3) If linearity As Found is ok, calibrate the instrument.
4) Check the linearity in As Left column.
5) In order to check linearity and perform calibration to other instrument, select
File New report on the pull down menu.