TelePole WR/ Operating & Maintenance Manual Page 15
Following is the operating procedure.
4.1.1. A short press on the B.CODE push-button turns the bar-code reader on. Three short
beeps will be sounded. The bAr. LCDs are displayed.
4.1.2. If the bar-code reading is successful, the Go. LCDs will be displayed.
4.1.3. To save the instrument data (ID, time, date, value, unit, comments) perform a short
press on the STORE push-button.
4.1.4. After the saving operation, the amount of available memory space remaining is
displayed.
4.1.5. To load data from the TelePole to the survey mapping software, connect the RS232 of
the PC to COM-1 on the TelePole rear panel by a special communication cable (supplied
with the software system). Downloading data from the TelePole to the PC clears the
instrument’s memory. TelePole can store up-to 380 readings (or 1550 optional with
extended memory).
Warning: To purposely clear data from the TelePole’s memory press the B.CODE push-
button for 10 seconds, the current available memory space is displayed for an additional
ten seconds. An additional press on the B.CODE push-button within the ten seconds erases
the data. The instrument indicates it by four audible beeps and 380 m, the new available
memory space is displayed.
4.2. WRM Communication
The WRM transmitter is connected to the serial port COM-2 located in the instrument’s rear
side via a standard telephone cable. The additional operations like transfer of data, are
performed automatically. The WRM communication protocol includes accumulated dose
data. Note: The TelePole does not measure accumulated dose, therefore instead of
transmitting the accumulated dose value, it transmits the digits 888888.8.
4.3. WRM2 Communication
The WRM2 optional kit(WR2-A??) is connected via two screws on the meter box and is
powered by the same power supply that powers the TelePole. The transmission interval
level is defaulted 4 seconds and can be changed using the Meter View software with the
communication cable designed for the TelePole and the Ram Ion and the baud rate to the
Telepole when using the WRM2 radio is set to 19,200 BPS. The WRM2 kit is available to
older instruments built 2006 or later and require new firmware and two wired jumpers
installed. Instruments built in 2007 or greater are pre-configured for the WRM2 Telepole
kit. Note: The TelePole does not measure accumulated dose, therefore instead of
transmitting the accumulated dose value, it transmits the digits 888888.8.