TelePole WR/ Operating & Maintenance Manual Page 53
4.7 Expose the TelePole to a radiation field of 600 R/h ±100 R/h. The meter will be
automatically set to factor F5. Now the factor can be changed either by pressing the
RESET or LIGHT push-button, or taking the meter outside the radiation chamber, the
radiation field should be lower than 10 mR/h. Notice if the current displayed radiation
factor is F5. If the factor is not F5 press the MODE push-button to switch to F5.
Adjust F5 by pressing the RESET or LIGHT push-button.
Place the TelePole back inside the radiation chamber and expose it to a radiation field
of 600 R/h. Check that the meter's reading and the radiation field are equal.
5. When adjusting F2, F3, F4, F5, and the TelePole is taken out of the radiation chamber, it
can either remain at the current factor or switch back to F1, depends on the radiation field
falling speed. If the meter switches to F1, press the MODE push-button to switch to the
current factor, each press will precede the factor to the next factor.