TelePole WR/ Operating & Maintenance Manual Page 19
Change the factor by pressing the RESET or LIGHT push-button to obtain the desired
reading.
5.2.3. Expose the detector to a radiation field of 100 mR/h to 800 mR/h. Note: Use calibration
points 800 mR/h to 1500 mR/h when using the optional switching 070301 firmware
(2000 mR/h – 2500 mR/h).
The display will show:
F2 F (factor) mR/h (reading)
_______________________________
Change the factor by pressing the RESET or LIGHT push-button to obtain the desired
reading.
5.2.4. Expose the detector to a radiation field of 10 R/h
±
±±
±
2 R/h.
The display will show:
F3 F (factor) R/h (reading)
_______________________________
Change the factor by pressing the RESET or LIGHT push-button to obtain the desired
reading.
5.2.5. Expose the detector to a radiation field of 200 R/h
±
±±
±
60 R/h.
The display will show:
F4 F (factor) R/h (reading)
_______________________________
Change the factor by pressing the RESET or LIGHT push-button to obtain the desired
reading.
5.2.6. Expose the detector to a radiation field of 500 R/h to 700 R/h.
The display will show:
F5 F (factor) R/h (reading)
_______________________________
Change the factor by pressing the RESET or LIGHT push-button to obtain the desired
reading.
5.2.7. Set the TelePole to the operating mode, turn the internal switch to the operating position.
5.3. CPU Unit Pulser Check (Option Only)
Note: Although it is possible to "calibrate" the TelePole using a series of input pulses (from a
pulser), the manufacturer strongly recommends against this. Checking instrument response
to input pulses ensures that the meter's CPU is operating properly, but indicates nothing
about the detector. The manufacturer recommends that the instrument is calibrated
according to the previous section (5.2 Calibration procedure), by exposing the probe head to
known radiation fields and adjusting factors appropriately, which ensures that both the
detectors and the CPU are operating properly.