TelePole WR/ Operating & Maintenance Manual Page 52
Appendix 2
Generic TelePole Calibration
1. It is possible to set/change all the factors outside the radiation chamber.
2. When the radiation field is lower than 10 mR/h, it is possible to change all the factors.
3. When in the calibration mode and the external radiation field is lower than 10 mR/h,
pressing the MODE push-button can switch between the different factors.
4. Suggested procedure:
4.1 Turn the internal switch to the calibration position.
4.2 Press the MODE push-button twice to enter the calibration mode.
4.3 Expose the TelePole to a radiation field of 50 mR/h ±20mR/h. The meter will be
automatically set to factor F1. Now the factor can be changed either by pressing the
RESET or LIGHT push-button or taking the meter outside the radiation chamber and
adjust the factor F1.
Place the TelePole back inside the radiation chamber and expose it to a radiation field
of 50 mR/h. Check that the meter's reading and the radiation field are equal.
4.4 Expose the TelePole to a radiation field of 350 mR/h ±50mR/h. The meter will be
automatically set to factor F2. Now the factor can be changed either by pressing the
RESET or LIGHT push-button or taking the meter outside the radiation chamber, the
radiation field should be lower than 10 mR/h. Notice if the current displayed radiation
factor is F2. If the factor is not F2 press the MODE push-button to switch to F2.
Adjust F2 by pressing the RESET or LIGHT push-button.
Place the TelePole back inside the radiation chamber and expose it to a radiation field
of 350 mR/h. Check that the meter's reading and the radiation field are equal.
4.5 Expose the TelePole to a radiation field of 10 R/h ±2 R/h. The meter will be
automatically set to factor F3. Now the factor can be changed either by pressing the
RESET or LIGHT push-button or taking the meter outside the radiation chamber, the
radiation field should be lower than 10 mR/h. Notice if the current displayed radiation
factor is F3. If the factor is not F3 press the MODE push-button to switch to F3.
Adjust F3 by pressing the RESET or LIGHT push-button.
Place the TelePole back inside the radiation chamber and expose it to a radiation field
of 10 R/h. Check that the meter's reading and the radiation field are equal.
4.6 Expose the TelePole to a radiation field of 350 R/h ±50 R/h. The meter will be
automatically set to factor F4. Now the factor can be changed either by pressing the
RESET or LIGHT push-button or taking the meter outside the radiation chamber, the
radiation field shold be lower than 10 mR/h. Notice if the current displayed radiation
factor is F4. If the factor is not F4 press the MODE push-button to switch to F4.
Adjust F4 by pressing the RESET or LIGHT push-button.
Place the TelePole back inside the radiation chamber and expose it to a radiation field
of 350 R/h. Check that the meter's reading and the radiation field are equal.